Remote control commands
R&S
®
RTO6
2159User Manual 1801.6687.02 ─ 05
TRIGger<m>:USB:SMAX <SEUT
o>
Sets the the end value of an SEU range if TRIGger<m>:
USB:SCONdition is set to
INRange or OORange.
Parameters:
<SEUTo>
TRIGger<m>:USB:SPEC <USBSpecialType>
Sets the trigger to one out of four dif
ferent Special PID types: PREamble, ERR, SPLit,
or PING.
Parameters:
<USBSpecialType> PREamble | ERR | SPLit | PING
PREamble
Sets the trigger to the PREamble PID
ERR
Sets the trigger to the ERRor PID
SPLit
Sets the trigger to the SPLIT PID (in USB high speed transac-
tions)
PING
Sets the trigger to the PING PID (in USB high speed transac-
tions, flow control probe for a bulk/control endpoint)
*RST: PREamble
TRIGger<m>:USB:STCO <SC>
Sets the trigger to a specific start-split or complete-split transaction endpoint.
Parameters:
<SC> ONE | ZERO | DC
ONE
SC = 1 represents a complete-split (CSPLIT) transaction.
ZERO
SC = 0 represents a start-split (SSPLIT) transaction.
DC
SC = X represents "don't care" (DC)
*RST: DC