Remote control commands
R&S
®
RTO6
2542User Manual 1801.6687.02 ─ 05
POWer:ONOFf:REPort:ADD.........................................................................................
2542
POW
er:ONOFf:RESult:TOFF?...................................................................................... 2543
POWer:ONOFf:RESult:TON?........................................................................................ 2543
POWer:ONOFf:TIME....................................................................................................2543
POWer:ONOFf:TYPE................................................................................................... 2543
POWer:ONOFf:ATOFf <ACTrigLevOff>
POWer:ONOFf:ATON <ACTrigLevOn>
Triggers the beginning of the measurements at the moment the AC input voltage rea-
ches the set value.
Parameters:
<ACTrigLevOn> Range:  -1E+6  to  1E+6
Increment:  1E-3
*RST:  10
Default unit: V
POWer:ONOFf:DTOFf <ACT
rigLevOf
f>
POWer:ONOFf:DTON <ACTrigLevOn>
POWer:ONOFf:DSOFf <StateLevelOff>
POWer:ONOFf:DSON <StateLevelOn>
Sets the percentage of the steady state level of the DC output that has to be reached.
Parameters:
<StateLevelOn> Range:  0  to  100
Increment:  1
*RST:  90
Default unit: %
POWer:ONOFf:EXECute
Starts the turn on/off measurement.
Usage:  Event
POWer:ONOFf:INPut <InputType>
Sets the input type.
Parameters:
<InputT
ype> AC | DC
*RST:  AC
POWer:ONOFf:REPort:ADD
Adds the result to the report list.
Usage:  Event
Power analysis (option R&S RT
O6-K31)