EasyManua.ls Logo

R&S RTO6 - Page 1371

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1371User Manual 1801.6687.02 ─ 05
<n> 1..4
Parameters:
<ShuntMode> INTShunt | EXTShunt
*RST: INTShunt
Usage: Asynchronous command
ZVC:Z<m>:I<n>:SHUNt:MAXCurrent <IntShuntMeasRg>
Sets the maximum current and the internal shunt value.
The internal shunt and maximum current parameter pairs are defined as described in
the table below.
Parameter Internal shunt Maximum current
A100r01
10 A 10 m
A40R01
4.5 A 10 m
A45M10r
45 mA 10
A4M510r
4.5 mA 10
A45u10k
45 µA 10 K
A4U510k
4.5 µA 10 K
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<IntShuntMeasRg> A100r01 | A40R01 | A45M10r | A4M510r | A45u10k | A4U510k
*RST: A100r01
Usage: Asynchronous command
ZVC:Z<m>:I<n>:SHUNt:MXCValue?
Queries the maximum current.
Suffix:
<m>
.
1..2
<n> 1..4
Return values:
<MaxCurrentValue> Default unit: A
Usage: Query only
Asynchronous command