Remote control commands
R&S
®
RTO6
1370User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<VertPosi> Range: -5 to 5
Increment: 0.02
*RST: 0
Default unit: div
Usage: Asynchronous command
ZVC:Z<m>:I<n>:SCALe <VerticalScale>
Sets the vertical scale for the current channel in V
olts per division. Increasing the scale
compresses the display of the signal.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<VerticalScale> Range: 1E-15 to 10E+24
Increment: 10E-6
*RST: 3
Default unit: V/div
Usage: Asynchronous command
ZVC:Z<m>:I<n>:SHUNt:EVALue <ExtShuntVal>
Defines the value of the external shunt resistor to calculate the correct current values.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<ExtShuntV
al> Range: 10E-6 to 10000
Increment: 0.02
*RST: 1
Default unit: Ω
Usage: Asynchronous command
ZVC:Z<m>:I<n>:SHUNt:MODE <ShuntMode>
Sets the internal or external shunt mode.
Suffix:
<m>
.
1..2