Remote control commands
R&S
®
RTO6
1553User Manual 1801.6687.02 ─ 05
Example: See: Chapter 24.3.2.2, "Setting reference levels", on page
1262
Usage: Asynchronous command
REFLevel<m>:LMODe
<Mode>
Defines whether the reference is configured using absolute or relative values.
Suffix:
<m>
.
Valid suf
fix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 24.4.1,
"Waveform suffix", on page 1279.
Parameters:
<Mode> ABS | REL
Example:
REFLevel2:LMODe ABS
Sets definition of reference levels to absolute values for C1W1.
C1W1 corresponds to suf
fix number 2.
Example: See: "Manual reference level definition using relative values"
on page 1262
Usage: Asynchronous command
REFLevel<m>:RELative:MODE <Mode>
The lower, middle and upper reference levels, defined as percentages of the high sig-
nal level.
Suffix:
<m>
.
Valid suf
fix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 24.4.1,
"Waveform suffix", on page 1279.
Parameters:
<Mode> FIVE | TEN | TWENty | USER
FIVE
5/50/95
TEN
10/50/90
TWENty
20/50/80
USER
Set the reference levels to individual values with
REFLevel<m>:
RELative:
LOWer, REFLevel<m>:
RELative:MIDDle, and REFLevel<m>:RELative:UPPer.
Automatic measurements