Remote control commands
R&S
®
RTO6
1560User Manual 1801.6687.02 ─ 05
Example:
REFLevel8:RELative:LOWer 85
Sets the upper reference level for C3W1 to 85 %. C3W1 corre-
sponds to suffix number 8.
Example: See:
"Manual reference level definition using relative values"
on page 1262
Usage: Asynchronous command
REFLevel<m>:RELative:MIDDle <Value>
Sets the middle relative reference level if REFLevel<m>:
RELative:MODE is set to
USER.
Suffix:
<m>
.
Valid suf
fix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 24.4.1,
"Waveform suffix", on page 1279.
Parameters:
<Value> Percentage of the high signal level.
Example:
REFLevel8:RELative:MIDDle 50
Sets the middle reference level for C3W1 to 50 %. C3W1 corre-
sponds to suf
fix number 8.
Example: See: "Manual reference level definition using relative values"
on page 1262
Usage: Asynchronous command
REFLevel<m>:RELative:LOWer <Value>
Sets the lower relative reference level if REFLevel<m>:
RELative:MODE is set to
USER.
Suffix:
<m>
.
Valid suf
fix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 24.4.1,
"Waveform suffix", on page 1279.
Parameters:
<Value> Percentage of the high signal level.
Example:
REFLevel8:RELative:LOWer 15
Sets the lower reference level for C3W1 to 15 %. C3W1 corre-
sponds to suf
fix number 8.
Example: See: "Manual reference level definition using relative values"
on page 1262
Usage: Asynchronous command
Automatic measurements