Remote control commands
R&S
®
RTO6
1559User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
V
alid suf
fix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 24.4.1,
"Waveform suffix", on page 1279.
Parameters:
<Value>
Example:
REFLevel2:ABSolute:ULEVel 0.01
Sets the upper reference level for C1W1 to 10 mV
. C1W1 corre-
sponds to suf
fix number 2.
Example: See: "Manual reference level definition using absolute values"
on page 1262
Usage: Asynchronous command
REFLevel<m>:ABSolute:LLEVel <Value>
The lower reference level, required e.g. to determine a fall.
Suffix:
<m>
.
Valid suf
fix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 24.4.1,
"Waveform suffix", on page 1279.
Parameters:
<Value>
Example:
REFLevel2:ABSolute:LLEVel 0.001
Sets the lower reference level for C1W1 to 1 mV
. C1W1 corre-
sponds to suffix number 2.
Example: See: "Manual reference level definition using absolute values"
on page 1262
Usage: Asynchronous command
REFLevel<m>:RELative:UPPer <Value>
Sets the upper relative reference level if REFLevel<m>:
RELative:MODE is set to
USER.
Suffix:
<m>
.
Valid suf
fix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 24.4.1,
"Waveform suffix", on page 1279.
Parameters:
<Value> Percentage of the high signal level.