Remote control commands
R&S
®
RTO6
1487User Manual 1801.6687.02 ─ 05
Parameters:
<Identifier> ON | OFF
*RST:  OFF
Example: 
:MEASurement:ARNames 1
:MEASurement:ARES?
Low: -1.185770750988e-001,Amplitude:
2.371541501976e-001,Max:
 1.343873517787e-001
Usage:  Asynchronous command
MEASurement<m>:RESult[:ACTual]? [<MeasType>]
MEASurement<m>:
RESult:AVG? [<MeasType>]
MEASurement<m>:RESult:EVTCount? [<MeasType>]
MEASurement<m>:RESult:NPEak? [<MeasType>]
MEASurement<m>:RESult:PPEak? [<MeasType>]
MEASurement<m>:RESult:RELiability? [<MeasType>]
MEASurement<m>:RESult:RMS? [<MeasType>]
MEASurement<m>:RESult:WFMCount? [<MeasType>]
MEASurement<m>:RESult:STDDev? [<MeasType>]
Return the statistic results of the specified measurement. If no parameter is specified,
the result of the main measurement is returned. The main measurement is defined
using MEASurement<m>:MAIN.
●
[:ACTual]: current measurement result
●
AVG: average of the long-term measurement results
●
EVTCount: number of measurement results in the long-term measurement
●
NPEak: negative peak value of the long-term measurement results
●
PPEak: positive peak value of the long-term measurement results
●
RELiability: reliability of the measurement result
●
RMS: RMS value of the long-term measurement results
●
STDDev: standard deviation of the long-term measurement results
For a detailed description of the results, see "Selection of the measurement group:
MEASurement<m>" on page 1479.
Automatic measurements