Remote control commands
R&S
®
RTO6
1486User Manual 1801.6687.02 ─ 05
MEASurement<m>:ARES?
Returns the results of all active measurements for the selected measurement group. If
statistics are enabled, the instrument returns also statistical results.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
Return values:
<sData> Result string
Example:
:MEASurement:
ARES?
2.351778656126e-001,
1.343873517787e-001,
-1.343873517787e-001
Example:
:MEASurement:
ARNames 1
:MEASurement:ARES?
Amplitude:
2.351778656126e-001,Max:
1.343873517787e-001,Min: -1.343873517787e-001
Example:
:MEASurement:ARNames 1
:MEASurement:STATistics:
ENABle 1
:MEASurement:ARES?
Amplitude:
2.312252964427e-001,2.608695652174e-001,
2.292490118577e-001,2.391822576775e-001,
2.392462568683e-001,5.533663458383e-003,12447,
12447,Max: 1.343873517787e-001,
1.343873517787e-001,1.324110671937e-001,
1.340297640714e-001, 1.340319233732e-001,
7.608360495310e-004,12447,12447, Min:
-1.324110671937e-001,-1.324110671937e-001,
-1.343873517787e-001,-1.340169449387e-001,
1.340191641476e-001,7.712812700636e-004,12447,
12447
Results: current, peak+, peak-, average, RMS,
standard deviation, event count, waveform count
Usage: Query only
Asynchronous command
MEASurement<m>:ARNames <Identifier>
Enables a prefix that indicates the measurement in the result string of the
MEASurement<m>:ARES? command.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
on page 1479.
Automatic measurements