EasyManua.ls Logo

R&S RTO6 - Page 1406

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1406User Manual 1801.6687.02 ─ 05
TRIGger<m>:SETHold:CSOurce:LEVel <ClockLevel>
Sets the voltage level for the clock signal. Both this command and TRIGger<m>:
SETHold:
CSOurce:
EDGE define the starting point for calculation of the setup and
hold time.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockLevel>
Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
TRIGger<m>:SETHold:CSOurce[:VALue] <ClockSource>
Selects the source of the clock signal.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockSource>
CHAN1 | CHANnel1 | CHAN2 | CHANnel2 | CHAN3 |
CHANnel3 | CHAN4 | CHANnel4
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
Input channel of the clock signal
*RST
: CHAN1
Usage: Asynchronous command
TRIGger<m>:SETHold:HTIMe <HoldTime>
Sets the minimum time
after the clock edge while the data signal must stay steady
above or below the data level.
The hold time can be negative. In this case, the setup time is always positive. The
setup/hold interval starts before the clock edge (setup time) and ends before the clock
edge (hold time). If you change the negative hold time, the setup time is adjusted by
the intrument.
Suffix:
<m>
.
1..3
Only 1 = A-trigger
, 2 | 3 = not available. Can be omitted.
Trigger

Table of Contents

Other manuals for R&S RTO6

Related product manuals