ACQuire:ARESet:TIME <EnvelopeT
imeout>
Defines the time after which the envelope and average evaluation restarts.
The setting is relevant if
ACQuire:ARESet:MODE is set to TIME.
Parameters:
<EnvelopeTimeout> Range: 0.1 to 10000
Increment: 0.01
*RST: 0.1
Default unit: s
Usage: Asynchronous command
ACQuire:ARESet:COUNt <NofWaveforms>
Defines the number of acquired waveforms after which the envelope and average eval-
uation restarts.
The setting is relevant if
ACQuire:ARESet:MODE is set to WFMS.
Parameters:
<NofWaveforms> Range: 2 to 16777215
Increment: 10
*RST: 1000
Usage: Asynchronous command
POWer:SLEWrate:GATE:STARt <T0>
POWer:SLEW
rate:GATE:STOP <T1>
Sets the value for the cursor.
Parameters:
<T1>
POWer:SLEWrate:REPort:ADD
Adds the result to the report list.
Usage: Event
POWer:SLEWrate:RESult:ACTual? <MeasType>
POW
er:SLEWrate:RESult:AVG? <MeasType>
POWer:SLEWrate:RESult:EVTCount? <MeasType>
POWer:SLEWrate:RESult:NPEak? <MeasType>
POWer:SLEWrate:RESult:PPEak? <MeasType>
POWer:SLEWrate:RESult:RMS? <MeasType>
Power analysis (option R&S RT
O6-K31)