Remote control commands
R&S
®
RTO6
1404User Manual 1801.6687.02 ─ 05
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: Lower = -0.1, upper =0.1
Default unit: V
Usage: Asynchronous command
TRIGger<m>:SLEW:RANGe <RangeMode>
Defines the time limit for the slew rate in relation to the upper or lower trigger level (see
TRIGger<m>:SLEW:
RATE on page 1404 and TRIGger<m>:SLEW:DELTa
on page 1405). The time measurement starts when the signal crosses the first trigger
level - the upper or lower level depending on the selected slope - and stops when the
signal crosses the second level.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Parameters:
<RangeMode> INSRange | OUTRange | LTHan | GTHan
INSRange
Triggers on pulses inside a given range. The range is defined by
the slew rate
±delta.
OUTRange
Triggers on pulses outside a given range. The range is defined
by the slew rate ±delta.
LTHan
Triggers on pulses shorter than the given slew rate.
GTHan
Triggers on pulses longer than the given slew rate.
*RST: GTHan
Usage: Asynchronous command
TRIGger<m>:SLEW:RATE <Time>
For the ranges "Within"
and "Outside", the slew rate defines the center of a range
which is defined by the limits "±Delta".
For the ranges "Shorter" and "Longer", the slew rate defines the maximum and mini-
mum slew rate limits, respectively. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope (see TRIGger<m>:
SLEW:SLOPe on page 1403).
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
Trigger