Remote control commands
R&S
®
RTO6
2530User Manual 1801.6687.02 ─ 05
POWer:INRush:EXECute
Starts the inrush current measurement.
Usage: Event
POWer:INRush:GATE<m>:STARt <StartT
ime>
POW
er:INRush:GATE<m>:STOP <StopTime>
Sets the measuring time for the selected gate.
Suffix:
<m>
.
*
Parameters:
<StopTime> Range: 0 to 10
Increment: 0
*RST: 100E-6
Default unit: s
POWer:INRush:GATE<m>:VALue <Value>
Returns the value of the inrush current.
Suffix:
<m>
.
*
Parameters:
<V
alue> Range: -1000 to 1000
Increment: 0
*RST: 0
Default unit: A
POWer:INRush:MAXCurrent <MaxExpCurr>
Sets the maximum expected current for the vertical scale.
Parameters:
<MaxExpCurr> Range: -1000 to 1000
Increment: 0
*RST: 10
Default unit: A
POWer:INRush:TRIGger <CurrentV
alue>
Sets the current value for the trigger
.
Parameters:
<CurrentValue> Range: -1000 to 1000
Increment: 0
*RST: 1
Default unit: A