I/Q software interface (Option R&S RTO6-K1
1)
R&S
®
RTO6
1061User Manual 1801.6687.02 ─ 05
1 = Signal component
2 = Input type/Input mode
3 = Sideband
4 = Carrier frequency
Enable
Enables the I/Q mode to record I/Q data.
The analyzing tools on the toolbar are deactivated, and the magnitude of the I/Q vector
is displayed.
Remote command:
IQ:STATe on page
2473
Input sample rate
Shows the sample rate of capturing the input data in I/Q mode: 10
Gsample/s.
Remote command:
IQ:INSRate on page 2473
Input type, Input mode
"Input type" sets the format of the input signal. "Input mode" selects the frequency
band of a complex input signal.
The R&S RTO6 I/Q software interface supports three formats of input signals.
Real RF signals
The input signal is down-converted, filtered and resampled to the sample rate
of the output I/Q signal.
One channel is required for each input signal. Thus up to four real signals can
be recorded in parallel.
Sideband settings: see "Sideband (real input)" on page 1063.
Complex I/Q baseband signals
I/Q sampling setup