EasyManua.ls Logo

R&S RTO6 - Page 1375

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1375User Manual 1801.6687.02 ─ 05
Parameters:
<VerticalScale> Range: 1E-15 to 10E+24
Increment: 10E-6
*RST: 3
Default unit: V/div
Usage: Asynchronous command
ZVC:Z<m>:V<n>:SKEW <DeskewOffset>
Sets the skew of
fset value for the voltage probe channel. This is a delay value, that is
known from the circuit specifics but cannot be compensated by the instrument auto-
matically.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<DeskewOffset> Range: -6E-6 to 6E-6
Increment: 200E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
ZVC:Z<m>:V<n>[:STATe] <State>
Enables the corresponding voltage channel of the probe.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
ZVC:Z<m>:ID:NAME?
Queries the name of the probe.
Suffix:
<m>
.
1..2
Return values:
<Name>
Usage: Query only
Asynchronous command
Acquisition and setup

Table of Contents

Other manuals for R&S RTO6