Remote control commands
R&S
®
RTO6
1387User Manual 1801.6687.02 ─ 05
HDEFinition:STATe <State>
Activates the high definition mode of the instrument.
Parameters:
<State> ON | OFF
ON: high definition mode, up to 16 bit digital resolution
OFF: normal oscilloscope mode
*RST: OFF
Usage: Asynchronous command
HDEFinition:BWIDth <Bandwidth>
Sets the filter bandwidth for the high definition mode.
Parameters:
<Bandwidth> Range: 10000 to max. 2 GHz, depending on the instru-
ment bandwidth.
Increment: 1000
*RST: 1E+6
Default unit: Hz
Usage: Asynchronous command
See "Bandwidth
"
on page 134 for bandwidth limits.
HDEFinition:RESolution?
Returns the resulting digital resolution in high definition mode.
Return values:
<Resolution> Range: 0 to 16
Increment: 0.1
*RST: 0
Default unit: bit
Usage: Query only
Asynchronous command
FORMat:BORDer <ByteOrder>
Sets the endianess.
The command is only relevant for raw data export in high definition mode (16 bit word
length).
Parameters:
<ByteOrder> LSBFirst | MSBFirst
LSB first: little endian, least significant byte first
MSB first: big endian, most significant byte first
*RST: LSBFirst
Acquisition and setup