Remote control commands
R&S
®
RTO6
1390User Manual 1801.6687.02 ─ 05
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
Input channels
EXT
ernanalog
External analog signal connected to the External T
rigger Input.
For this source, only the analog edge trigger is available.
SBUS
Serial bus
D0...D15
Digital channels (option R&S RTO6-B1)
See also: Chapter 24.18.4, "Trigger settings for digital signals
and parallel buses", on page 2431
LOGIc
Logic combination of digital channels, used as trigger source
(option R&S RTO6-B1)
MSOB1 | MSOB2 | MSOB3 | MSOB4
Parallel bus (option R&S RTO6-B1)
Z1V1 | Z1V2 | Z1V3 | Z1V4 | Z1I1 | Z1I2 | Z1I3 | Z1I4 | Z2V1 |
Z2V2 | Z2V3 | Z2V4 | Z2I1 | Z2I2 | Z2I3 | Z2I4
Input channels of R&S RT-ZVCmulti-channel power probe. Only
available in the A-trigger with trigger type EDGE.
DIFF1 | DIFF2 | COMMON1 | COMMON2
Differential signals
*RST: CHAN1
Usage: Asynchronous command
TRIGger<m>:TYPE <Type>
Selects the trigger type to trigger on analog channels or the external trigger input.
See also: Chapter 6.3, "Trigger types", on page 193.
To trigger on digital channels and parallel buses, use TRIGger<m>:PARallel:TYPE.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
For suffixes 2 and 3, the following trigger types are availabe:
EDGE, GLITch, WIDTh, RUNT, WINDow, TIMeout, INTerval,
SLEWrate.
Parameters:
<Type> EDGE | GLITch | WIDTh | RUNT | WINDow | TIMeout | INTerval |
SLEW
rate |
STATe | PATTern | ANEDge | SERPattern | TV | |
SETHold | | NFC | CDR
Most of the type values are self-explanatory.
Trigger

Table of Contents

Other manuals for R&S RTO6

Related product manuals