Remote control commands
R&S
®
RTO6
1430User Manual 1801.6687.02 ─ 05
TRIGger<m>:MODE <T
riggerMode>
Sets the trigger mode which determines the behaviour of the instrument if no trigger
occurs.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<T
riggerMode> AUTO | NORMal | FREerun
AUTO
The instrument triggers repeatedly after a time interval if the trig-
ger conditions are not fulfilled. If a real trigger occurs, it takes
precedence. The time interval depends on the time base.
NORMal
The instrument acquires a waveform only if a trigger occurs.
FREerun
The instrument triggers after a very short time interval - faster
than in AUT
O mode. Real triggers are ignored.
*RST: AUTO
Usage: Asynchronous command
TRIGger<m>:FORCe
If the acquisition is running in normal mode and no valid trigger occurs, forcing the trig-
ger provokes an immediate single acquisition. Thus you can confirm that a signal is
available and use the waveform display to determine how to trigger on it.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Usage: Event
Asynchronous command
TRIGger<m>:OUT:POLarity <Polarity>
Sets the polarity of the trigger out pulse.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Polarity> POSitive | NEGative
*RST
: POSitive
Usage: Asynchronous command