Remote control commands
R&S
®
RTO6
1432User Manual 1801.6687.02 ─ 05
Parameters:
<Beep> NOACtion | TRIGger
*RST: NOACtion
Usage: Asynchronous command
TRIGger<m>:EVENt:PRINt <Print>
Saves a screenshot at each trigger if the command is set to TRIGger.
For screenshot settings, see Chapter 24.16.8, "Screenshots", on page 1660.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<Print> NOACtion | TRIGger
*RST
: NOACtion
Usage: Asynchronous command
TRIGger<m>:EVENt:WFMSave <SaveWfm>
Saves the waveform data to file at each trigger if the command is set to TRIGger.
For data export settings, see Chapter 24.16.4, "Waveform data export to file",
on page
1649
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<SaveWfm> NOACtion | TRIGger
*RST: NOACtion
Usage: Asynchronous command
TRIGger<m>:EVENt:RUNexec <RunExecutable>
Starts an external application if the command is set to TRIGger.
Use the following commands to set up the application:
●
EXECutable:NAME
on page 1668
●
EXECutable:PARameter on page 1669
●
EXECutable:WDIRectory on page 1669
Suffix:
<m>
.
1..3
The suf
fix is irrelevant.
Trigger