Remote control commands
R&S
®
RTO6
1495User Manual 1801.6687.02 ─ 05
MEASurement<m>:AMPTime:DELay<n>:ECOunt <EdgeIndex>
Sets the number of the edge that is relevant for delay measurement for each source.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page
1479.
<n> 1..2
Selects the source number.
Parameters:
<EdgeIndex>
Edge number
Range: 1 to 100000
Increment: 1
*RST: 1
Usage: Asynchronous command
MEASurement<m>:AMPTime:DELay<n>:LSELect <DlyLevSel>
Selects the reference level on which the time is measured for each source.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
<n> 1..2
Selects the source number.
Parameters:
<DlyLevSel>
UPPer | MIDDle | LOWer
*RST: MIDDle
Usage: Asynchronous command
MEASurement<m>:AMPTime:DELay<n>:SLOPe <Slope>
Sets the edge of each source, between which the delay is measured.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1479.
<n> 1..2
Selects the source number.
Parameters:
<Slope> POSitive | NEGative |
EITHer
*RST: POSitive
Automatic measurements