Remote control commands
R&S
®
RTO6
1585User Manual 1801.6687.02 ─ 05
Parameters:
<TolViolCnt> Range: 0 to 4000000000
Increment: 1
*RST: 0
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:CTYPe <MaskTestName>,<DefinitionType>
MTESt:CTYPe? <MaskTestName>
Sets the method of mask definition.
Parameters:
<DefinitionType> USER | WFML | EYEMask | PROTocol
USER
The mask segments are created by entering the numerical x-
and y-values of the mask points.
See: Chapter 24.14.2, "Mask definition: user mask",
on page 1586
WFML
The mask is created from the envelope of an existing waveform.
See:
Chapter 24.14.3, "Mask definition: waveform mask",
on page 1592
EYEMask
Mask for eye diagram testing. See: Chapter 24.22.6, "Eye mask
testing", on page 2519.
Requires jitter option R&S RTO6-K12
*RST: USER
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:FILE:NAME <MaskTestName>, <Path>
MTESt:FILE:
NAME? <MaskTestName>
Specifies a file to save the mask test.
Parameters:
<Path> String containing path and file name, format .xml
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Mask testing