Remote control commands
R&S
®
RTO6
1605User Manual 1801.6687.02 ─ 05
SEARch:REMove <Key>
Deletes the specified search definition.
Setting parameters:
<Key> String with the name of the search
Usage: Setting only
Asynchronous command
SEARch:SOURce <SearchName>,<Source>
SEARch:SOURce? <SearchName>
Defines the source on which the search conditions are applied. The source can be any
analog or digital channel, math or reference waveform as well as a serial bus config-
ured for a supported protocol.
Parameters:
<Source> C1W1 | C1W2 | C1W3 | C2W1 | C2W2 | C2W3 | C3W1 | C3W2 |
C3W3 | C4W1 | C4W2 | C4W3 | M1 | M2 |
M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | SBUS1 |
SBUS2 | SBUS3 | SBUS4
Source of the search, see Chapter 24.4.2, "Waveform parame-
ter", on page 1284
*RST: C1W1
Parameters for setting and query:
<SearchName> String with the name of the search
Example: See Chapter 24.3.4.1, "Searching for a pulse of specified width",
on page
1266
Usage: Asynchronous command
SEARch:ONLine <SearchName>,<OnlineState>
SEARch:ONLine? <SearchName>
If enabled, a search is performed repeatedly for each new data acquisition.
Parameters:
<OnlineState> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName> Search definition
Usage: Asynchronous command
SEARch:ALL <SearchName>
Performs a search for all results on the existing data from the selected source.