EasyManua.ls Logo

R&S RTO6 - Page 1612

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1612User Manual 1801.6687.02 ─ 05
Parameters:
<WidthDelta> Range: 100E-12 to 864
Increment: 100E-9
*RST: 100E-12
Default unit: s
Parameters for setting and query:
<SearchName> Search definition
Usage: Asynchronous command
SEARch:TRIGger:RUNT:POLarity <SearchName>,<Polarity>
SEARch:TRIGger:RUNT:
POLarity? <SearchName>
Indicates the polarity of a pulse, that is the direction of the first pulse slope.
Parameters:
<Polarity> POSitive | NEGative | EITHer
See Chapter 24.4.4, "Polarity parameter", on page 1285.
*RST
: POSitive
Parameters for setting and query:
<SearchName> Search definition
Usage: Asynchronous command
SEARch:TRIGger:RUNT:RANGe <SearchName>,<Mode>
SEARch:TRIGger:RUNT:
RANGe? <SearchName>
Selects how the time limit of the runt pulse is defined based on the runt width and delta
(see SEARch:TRIGger:RUNT:WIDTh on page 1613 and SEARch:TRIGger:RUNT:
DELTa on page 1611).
Parameters:
<Mode> ANY | LONGer | SHORter | WITHin | OUTSide
ANY
Triggers on all runts fulfilling the level condition, without time lim-
itation.
LONGer
T
riggers on runts longer than the given "Runt width".
SHORter
Triggers on runts shorter than the given "Runt width".
WITHin
Triggers if the runt length is inside a given time range. The range
is defined by "Runt width" and "±Delta".
OUTSide
Triggers if the runt length is outside a given time range. The
range definition is the same as for "Within" range.
*RST: ANY
Search

Table of Contents

Other manuals for R&S RTO6