Remote control commands
R&S
®
RTO6
1770User Manual 1801.6687.02 ─ 05
TRIGger<m>:CMSB:IMAX <DataPositionT
o>
Sets the end value of the bit index range (data position range), if TRIGger<m>:
CMSB:
ICONdition on page 1769 is set to "INRange".
Suffix:
<m>
.
1..3
Parameters:
<DataPositionTo> Range: 0 to 65535
Increment: 1
*RST: 65535
TRIGger<m>:CMSB:NRZ:WRDLength <NRZW
ordLength>
Sets the number of bits in an NRZ Unclocked word.
Suffix:
<m>
.
1..3
Parameters:
<NRZW
ordLength> Range: 0 to 31
Increment: 1
*RST: 8
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:ENABle <Enable>
TRIGger<m>:CMSB:ADVanced:
FRENable <Frame>, <Enabler>
Enables or disables the specific frame to be triggered on.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Setting parameters:
<Frame>
<Enabler> ON |
OFF
Example:
:TRIGger:CMSB:ADVanced:FRENable "Master" 1
Enables the Master frame.
:TRIGger:
CMSB:ADVanced:
FRAMe1:ENABle 1
Enables the 1st frame that you have defined.
Usage: Setting only
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:ENABle <CondEnabler>
TRIGger<m>:CMSB:ADVanced:
FIENable <Frame>, <Field>, <Enabler>
Enables or disables the specific field within the defined frame to be triggered on.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Protocols