Remote control commands
R&S
®
RTO6
1796User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<DataSource> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 |
R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
*RST: C1W1
Usage: Asynchronous command
BUS<m>:CAN:TYPE <SignalT
ype>
Selects the CAN-High or CAN-Low line. Both lines are required for dif
ferential signal
transmission used by CAN.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SignalType> CANH | CANL
*RST: CANL
BUS<m>:CAN:FDATa:PSTandard <ProtStd>
Only available for CAN FD buses. Selects whether the tested signal is an ISO CAN FD
signal or not.
Suffix:
<m>
.
1..4
Parameters:
<ProtStd> ISO | NISO
ISO
Signals are decoded according to the the ISO CAN FD protocol.
This protocol has an additional stuff count field before the CRC
sequence.
NISO
Non-ISO. Signals are decoded according to the the Bosch CAN
FD protocol.
*RST
: ISO
Firmware/software: FW 3.35
BUS<m>:CAN:DATA:THReshold <Threshold>
Sets a user-defined threshold value.
Protocols