Remote control commands
R&S
®
RTO6
1843User Manual 1801.6687.02 ─ 05
For complete definition, set also the operator with TRIGger<m>:
LIN:DLECondition
on page
1842.
Parameters:
<WordCount> Range: 1 to 8
Increment: 1
*RST: 1
TRIGger<m>:LIN:IDOR<n>:ENABle <UseIdentifier>
Includes the indicated IDOR address in the "identifier OR" trigger condition.
Suffix:
<n>
.
1..4
Index of the identifier in an "identifier OR" condition
Parameters:
<UseIdentifier> ON | OFF
*RST: OFF
TRIGger<m>:LIN:IDOR<n>[:VALue] <IdPattern>
Defines the pattern of the indicated IDOR identifier in the "identifier OR" trigger condi-
tion.
Suffix:
<n>
.
1..4
Index of the identifier in an "identifier OR" condition
Parameters:
<IdPattern> Numeric or string pattern, see Chapter 24.4.6, "Bit pattern
parameter", on page
1286. The parameter accepts the bit value
X (don't care).
TRIGger<m>:LIN:SYERror <SyncError>
Triggers if a synchronization error occurs.
The trigger type has to be set before:
TRIGger<m>:LIN:TYPE to ERRC.
Parameters:
<SyncError> ON | OFF
*RST: ON
TRIGger<m>:LIN:IPERror <IdParityError>
Triggers if an error occurs in the identifier parity bits. These are the bits 6 and 7 of the
identifier
.
The trigger type has to be set before: TRIGger<m>:LIN:TYPE to ERRC.
Protocols