EasyManua.ls Logo

R&S RTO6 - Page 1915

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1915User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the number of the frame in the current acquisition, 1...n.
Return values:
<Translation> String with symbolic label of the identifier.
Example:
BUS:SENT:FRAMe:SYMBol?
Response: Air Temperature
Usage: Query only
BUS<m>:SENT:FRAMe<n>:SYNCduration?
Returns the time of the synchronization pulse.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<SyncDuration> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:SENT:FRAMe<n>:TYPE?
Returns the type of SENT message.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<FrameType> TRSQ | SMSG | EMSG
T
ransmission sequence, short serial message or enhanced
serial message.
*RST
: TRSQ
Usage: Query only
Protocols