Remote control commands
R&S
®
RTO6
2000User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Parameters:
<SyncBitPos> Range: 0 to 10
Increment: 1
*RST: 0
BUS<m>:SWIRe:DATA:HYSTeresis <HysteresisData>
Sets a value for the hysteresis of the data signal.
Suffix:
<m>
.
1..4
Parameters:
<HysteresisData> Range: 0 to 0.8
Increment: 0.01
*RST: 0.2
Default unit: V
BUS<m>:SWIRe:DATA:SOURce <SourceData>
Selects the source of the data signal.
Suffix:
<m>
.
1..4
Parameters:
<SourceData> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 |
R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
*RST: C1W1
BUS<m>:SWIRe:DATA:THReshold <ThresholdData>
Sets the threshold value for the digitization of the data signal.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdData> Range: -10 to 10
Increment: 0.1
*RST: 1.25
Default unit: V
BUS<m>:SWIRe:MGAP <MinGapTime>
Sets the minimum duration of a gap. Any inactivity greater than this time will be inter-
preted as a gap and lead to a resynchronization to the signal.
Protocols

Table of Contents