EasyManua.ls Logo

R&S RTO6 - Page 2028

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
2028User Manual 1801.6687.02 ─ 05
Return values:
<Type> Returns the word type (specific for the sub-protocol), or the
length of the selected word.
Range: 0 to 65535
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:ETHernet:WORD<n>:DATA?
Returns the number of word bytes in brackets [.] followed by the first six word bytes of
data in hexadecimal format.
Use BUS<m>:ETHernet:WORD<n>:BYTE<o>:VALue? to access the word bytes.
Suffix:
<m>
.
1..4
<n> *
Return values:
<Data> String parameter
Example:
BUS:ETHernet:WORD3:DATA?
<-- '[60]FF00FFFF1234'
Usage: Query only
BUS<m>:ETHernet:WORD<n>:CRC?
Returns the Cyclic Redundancy Code (CRC, or frame check) checksum of the selected
word.
Suffix:
<m>
.
1..4
<n> *
Return values:
<CRC> Range: 0 to 4294967295
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:ETHernet:WORD<n>:DSYMbol?
Returns the symbolic label (or translation) of the destination address of the specified
word, if the label list is enabled.
Suffix:
<m>
.
1..4
<n> *
Protocols

Table of Contents

Other manuals for R&S RTO6

Related product manuals