Remote control commands
R&S
®
RTO6
2046User Manual 1801.6687.02 ─ 05
Usage:  Query only
BUS<m>:MDIO:WORD<n>:PHYS?
Returns the 5-bit address field content (PHY
AD/PR
TAD) in the present acquisition of
the selected word and the selected serial bus.
Suffix: 
<m>
.
1..4
<n> *
Return values: 
<PhyAd> Range:  0  to  32
Increment:  1
*RST:  0
Usage:  Query only
BUS<m>:MDIO:WORD<n>:REGI?
Returns the 5-bit register or device address field content (REGAD/DEVAD) in the pres-
ent acquisition of the selected word and the selected serial bus.
Suffix: 
<m>
.
1..4
<n> *
Return values: 
<RegAd> Range: 
0  to  32
Increment:  1
*RST:  0
Usage:  Query only
BUS<m>:MDIO:WORD<n>:ST?
Returns the Start Code (= start of frame code) in the present acquisition of the selected
word and the selected serial bus.
Suffix: 
<m>
.
1..4
<n> *
Return values: 
<StartCode> Range:  0  to  3
Increment:  1
*RST:  0
The parameter value "0" represents Clause 45, and "1" stands
for Clause 22.
The values "2"
 
and "3" do not correspond with any legal parame-
ters according to the standard, but they can be searched for.
Protocols