Remote control commands
R&S
®
RTO6
2054User Manual 1801.6687.02 ─ 05
SEARch:RESult:MDIO:WORD<m>:STOP? <SearchName>
Returns the stop time of the frame from the selected word within the search result.
Suffix:
<m>
.
*
Query parameters:
<SearchName>
Return values:
<Stop> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
SEARch:RESult:MDIO:WORD<m>:SYMBol? <SearchName>
Returns a textual translation (called Register Name) of the PHY or port address label
from the selected word within the search result.
Suffix:
<m>
.
*
Query parameters:
<SearchName>
Return values:
<Translation>
Usage: Query only
SEARch:RESult:MDIO:WORD<m>:TYPE? <SearchName>
Returns the frame type (= operation code or OpCode) for the selected word within the
search result.
Suffix:
<m>
.
*
Query parameters:
<SearchName>
Return values:
<FrameType> Range: 0 to 3
Increment: 1
*RST: 0
Note that the user interface shows interpretations of the numeri-
cal OpCode values corresponding to Clause 45: "0" (= [bin]00)
represents Address, "1" (= [bin]01" represents W
rite, "2" (=
[bin]10) represents Post Read, "3" (= [bin]1
1) represents Read.
Clause 22 is not represented by this interpretation.