Remote control commands
R&S
®
RTO6
2058User Manual 1801.6687.02 ─ 05
Parameters:
<ThresClkPos> Range: 0 to 2.5
Increment: 0.1
*RST: 0.72
Default unit: V
BUS<m>:RFFE:CLOCk:THReshold:LOW <ThresClkNeg>
Set the negative going threshold (V
TN
) for the clock signal.
Note that the high and low thresholds and the hysteresis values are dependent, so it is
suf
ficient to define two values for each line.
Suffix:
<m>
.
1..4
Parameters:
<ThresClkNeg>
Range: 0 to 2.5
Increment: 0.1
*RST: 0.48
Default unit: V
BUS<m>:RFFE:CLOCk:THReshold:HYSTeresis <ThresClkHyst>
Set the hysteresis between V
TP
and V
TN
for the clock signal.
Note that the high and low thresholds and the hysteresis values are dependent, so it is
suf
ficient to define two values for each line.
Suffix:
<m>
.
1..4
Parameters:
<ThresClkHyst>
Range: 0 to 1
Increment: 0.1
*RST: 0.2
Default unit: V
BUS<m>:RFFE:GFILter <GlitchFilter>
Enables the glitch filter on the SCLK and SDAT
A lines to improve decode accuracy.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<GlitchFilter> ON | OFF
*RST: OFF
Protocols