Triggers
R&S
®
RTO6
206User Manual 1801.6687.02 ─ 05
Slope
Sets the edge type for the trigger condition.
"Positive"
Selects the rising edge, that is a positive voltage change.
"Negative"
Selects the falling edge, that is a negative voltage change.
"Both"
Selects the rising as well as the falling edge. This option is not availa-
ble if the trigger source is the external trigger input.
Remote command:
TRIGger<m>:EDGE:
SLOPe on page 1392
TRIGger<m>:ANEDge:SLOPe on page 1436
TRIGger<m>:SLEW:SLOPe on page 1403
Upper level
Sets the upper voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
Remote command:
TRIGger<m>:LEVel<n>:SLEW:UPPer on page 1403
Lower level
Sets the lower voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
Remote command:
TRIGger<m>:LEVel<n>:SLEW:LOWer on page 1403
Range
Selects how the time limit for the slew rate is defined. The time measurement starts
when the signal crosses the first trigger level - the upper or lower level depending on
the selected slope. The measurement stops when the signal crosses the second level.
Trigger types