Remote control commands
R&S
®
RTO6
2123User Manual 1801.6687.02 ─ 05
<n> *
Selects the frame.
Return values:
<SrcTranslation>
Usage: Query only
BUS<m>:TBTO:RESult:FRAMe<n>:STARt?
Returns the start time of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Start> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:TBTO:RESult:FRAMe<n>:STATe?
Returns the state of the frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<State> OK | ERRFEC | ERRZERO | ERROOR | ERRCRC | INComplete
OK
V
alid frame.
ERRFEC
Reed-Solomon forward error correction.
ERRZERO
Zero ternary symbol pair error
.
ERROOR
Values are out of range error.
ERR_CRC
Erroneous frame due to CRC error.
INComplete
Incomplete frame.