Remote control commands
R&S
®
RTO6
2239User Manual 1801.6687.02 ─ 05
SEARch:RESult:USBPd:FRAMe<m>:FLD<n>:STATus? <SearchName>
Returns the field status of the field with the selected field number within the frame with
the selected frame number within the search result.
Suffix:
<m>
.
*
<n> *
Selects the field number.
Query parameters:
<SearchName>
Return values:
<FieldState>
OK | UNKNown | INComplete | PREamble | SOP | CRC |
LENGth | FBFB
SOP
Start of packet warning
CRC
Cyclic redundancy check error
FBFB
4b5b error
*RST: OK
Usage: Query only
SEARch:RESult:USBPd:FRAMe<m>:FLD<n>:STARt? <SearchName>
Returns the start time of the field with the selected field number within the frame with
the selected frame number within the search result.
Suffix:
<m>
.
*
<n> *
Query parameters:
<SearchName>
Return values:
<Start> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
SEARch:RESult:USBPd:FRAMe<m>:FLD<n>:STOP? <SearchName>
Returns the stop time of the field with the selected field number within the frame with
the selected frame number within the search result.