Remote control commands
R&S
®
RTO6
2259User Manual 1801.6687.02 ─ 05
<n> *
Selects the frame.
Return values:
<State> OK | CS | ECC | ECCWarning | LENGth | FORMat |
UNKNOWN_FRAME | INComplete
OK
The frame is valid.
LENGth
Length error.
UNKNOWN_FRAME
Unknown frame type
INComplete
The sequence is not completely contained in the acquisition
*RST
: OK
Usage: Query only
BUS<m>:DPHY:RESult:FRAMe<n>:STOP?
Returns the end time of the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Stop> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:DPHY:RESult:FRAMe<n>:TYPE?
Returns the type of frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<FrameType> HS_SP | HS_LP | HS_INCOMPLETE | LP_HSREQ | LP_TA |