Triggers
R&S
®
RTO6
238User Manual 1801.6687.02 ─ 05
"RF reject"
Frequencies higher than the "Cut-off" frequency are rejected, lower
frequencies pass the filter
. Y
ou can adjust the "Cut-off" frequency, the
default is 50 kHz.
Remote command:
TRIGger<m>:ANEDge:
FILTer on page 1434
TRIGger<m>:ANEDge:CUToff:HIGHpass on page 1435
TRIGger<m>:ANEDge:CUToff:LOWPass on page 1435
Slope
Sets the edge type for the trigger condition.
"Positive"
Selects the rising edge, that is a positive voltage change.
"Negative"
Selects the falling edge, that is a negative voltage change.
"Both"
Selects the rising as well as the falling edge. This option is not availa-
ble if the trigger source is the external trigger input.
Remote command:
TRIGger<m>:EDGE:
SLOPe on page 1392
TRIGger<m>:ANEDge:SLOPe on page 1436
TRIGger<m>:SLEW:SLOPe on page 1403
Trigger level
Sets the voltage level for the trigger condition. You can also drag the trigger level
marker on the display (TA or TB on the right edge of the display). The range of the trig-
ger level is limited in a way so that always a hysteresis for stable trigger conditions is
available.
Remote command:
TRIGger<m>:LEVel<n>[:VALue] on page 1391
External trigger input