Measurements
R&S
®
RTO6
298User Manual 1801.6687.02 ─ 05
3. Select the measurement "Category", for example, "Amp/Time".
4.
T
ap "Source" and select the waveform to be measured.
Spectrum measurements require an FFT math waveform as measurement source.
Histogram measurements require a histogram as source.
5. Tap "Add / Remove" and select all measurements that you want to include in the
measurement group. Tap "Clear" to deselect all settings.
6. Tap "Back" to close the selection box.
All selected measurements are displayed in the list of active measurements.
7. A "Settings" icon indicates whether further settings are required. Tap the measure-
ment. A dialog box opens, and you can adjust the measurement.
The settings are explained in the following chapters:
Chapter 8.2.5.2, "Settings for amplitude/time measurements",
on page 321
Chapter 8.2.6.2, "Settings for eye diagram measurements", on page 333
Chapter 8.2.7.2, "Settings for spectrum measurements", on page 336
Chapter 8.2.8.5, "Settings for histogram measurement", on page 346
8. Optionally, define a gate to restrict the measurement to a part of the waveform. See
Chapter 8.2.3.1, "Using measurement gates", on page 304.
If you enabled the measurement with the toolbar icon and drew a rectangle on the
diagram, the gate is already defined and enabled.
Automatic measurements