Measurements
R&S
®
RTO6
319User Manual 1801.6687.02 ─ 05
Meas. type Symbol Description/Result
Pos. pulse T
PosPulse
Width of a positive pulse: time between a rising edge and the
following falling edge measured on the middle reference level.
The measurement requires at least one complete period of a
triggered signal. Measurement all events in the acquisition is
possible.
Neg. pulse T
NegPulse
Width of a negative pulse: time between a falling edge and the
following rising edge measured on the middle reference level.
The measurement requires at least one complete period of a
triggered signal. Measurement all events in the acquisition is
possible.
Period T
Period
Time between two consecutive waveform edges of the same
direction, measured on the middle reference level. The mea-
surement requires at least one complete period of a triggered
signal. Measurement all events in the acquisition is possible.
f
Frequency f
Period
Frequency of the signal, reciprocal value of the period.
f
Period
= 1 / T
Period
Pos. duty cycle R
PosCyc
Positive duty cycle: Width of a positive pulse in relation to the
period in %. The measurement requires at least one complete
period of a triggered signal. Multiple measurements are possi-
ble.
%100
Period
PosPulse
PosCyc
T
T
R
Neg. duty cycle R
NegCyc
Negative duty cycle: Width of a negative pulse in relation to the
period in %. The measurement requires at least one complete
period of a triggered signal. Multiple measurements are possi-
ble.
%100
Period
NegPulse
NegCyc
T
T
R
Delay
The measurement result is negative if the edge of the second
source comes before the edge of the first source.
See: "Settings for delay and phase measurements (analog sour-
ces)" on page 323
Phase
Phase difference between two waveforms.
Phase = Delay / Period * 360
Burst width
Duration of one burst, measured from the first edge to the last
Automatic measurements