431User Manual 1801.6687.02 ─ 05
Clock source, Clock edge, Clock level
Set the clock settings. Both "Clock level" and "Clock edge"
define the starting point for
calculation of the setup and hold time.
Remote command:
SEARch:TRIGger:SETHold:CSOurce on page 1622
SEARch:TRIGger:SETHold:CEDGe on page 1621
SEARch:TRIGger:SETHold:CLEVel on page 1622
Data level
Sets the voltage level for the data signal. At this level, the setup and hold time are
measured.
Remote command:
SEARch:TRIGger:LEVel[:VALue] on page 1607
Level
Sets the voltage level for trigger search. The value is used for all search event types
that require one trigger level. The search trigger level is search-specific, you can define
different levels in dif
ferent searches for the same event.
Remote command:
SEARch:TRIGger:LEVel[:VALue] on page 1607
Setup time, Hold time
Sets the minimum time before (Setup) and after (Hold) the clock edge while the data
signal must stay steady above or below the data level.
See also: "Setup time" on page 209 and "Hold time" on page 209.
Remote command:
SEARch:TRIGger:SETHold:STIMe on page 1623
SEARch:TRIGger:SETHold:HTIMe on page 1622
Search setup