Protocol analysis
R&S
®
RTO6
546User Manual 1801.6687.02 ─ 05
3. Select the "Source" of the measurement: "Serial bus". Select the bus configured for
the audio signal.
4.
Enable the measurement.
5.
Tap "Add / Remove " and select "Audio".
6. Select the "Audio channel" for which you want to analyze the trend.
7. In the "Measurements" dialog, select the "Plot" tab.
8. Under "Long term", select "Enable".
An empty trend diagram is displayed.
9. Tap "Settings".
10. In the "Measurement: Long Term Setup", set the "Horizontal scaling" to "Measure-
ments".
11. Set the number of "Measurements/point" to 1.
12. Enable "Continuous auto scale"
.
13. Start continuous acquisition.
The trend fills up with measurement points from left to right.
Audio signals (option R&S RTO6-K510)