EasyManua.ls Logo

R&S RTO6 - Page 1092

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Jitter analysis and clock data recovery
R&S
®
RTO6
1092User Manual 1801.6687.02 ─ 05
"Track": the measurement results over time for an acquired waveform. It is the
graphical interpretation of all measurement values of a single acquisition. Unlike
the histogram, it reveals trends of change in the analysis and preserves the timing
relationship of the measurement results to the signal.
"Spectrum": displays how the power of the signal is distributed over the frequency.
Remote command:
ADVJitter<m>:
RESult:
COMPonents<n>:HISTogram on page 2495
ADVJitter<m>:RESult:COMPonents<n>:TRACk on page 2496
ADVJitter<m>:RESult:COMPonents<n>:SPECtrum on page 2496
Unit
In the jitter results dialog, you can select the unit for the displayed results. The follow-
ing values are available:
For jitter measurements, you can select between absolute (seconds) and relative
(unit intervals) values.
For noise measurements, you can select between absolute (V) and relative (%)
values.
Remote command:
ADVJitter<m>:RESult:UNIT on page 2496
Advanced jitter and noise (option R&S R
T
O6- K133/K134)

Table of Contents

Other manuals for R&S RTO6

Related product manuals