Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1126User Manual 1801.6687.02 ─ 05
Current channel skew
Shows the skew settings of the channel connected to the current probe. Skew settings
are defined in the "Horizontal > Skew"
dialog box.
See also: Chapter 5.3.1, "Skew", on page 123
Overwrite present skew setup
If disabled, the instrument only stores the result of the auto deskew procedure as a
separate value and does not use it. This value can be used later for power measure-
ments. The general skew offset under "Current channel skew" remains unchanged.
If enabled, the result of the auto deskew procedure is used for all measurements on
the selected channel. It is shown under "Current channel skew".
Remote command:
POWer:DESKew:RESet on page 2521
Auto deskew result
Available only if "Overwrite present skew setup" is disabled.
"Deskew
value"
Result of the auto deskew.
Remote command:
POWer:
DESKew:TIME? on page 2521
Power measurement selection