Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1145User Manual 1801.6687.02 ─ 05
Maximum current
Sets the maximum expected current for the vertical scale. Set the value according to
your signal to avoid saturation.
Remote command:
POWer:INRush:MAXCurrent
on page 2530
Trigger current value (T)
Sets the current value for the trigger. The measurement starts after the signal of the
DUT reaches this current value.
Remote command:
POWer:INRush:TRIGger on page 2530
Gate Configuration
In this table, you can configure different gates (time periods). You can configure up to
five different gates. The time periods of the defined gates may overlap.
To add a gate press "Insert" or"Append". To remove a gate press "Remove".
Remote command:
POWer:INRush:ADD on page 2529
POWer:INRush:INSert on page 2529
POWer:INRush:REMove on page 2529
Inrush current ← Gate Configuration
Shows the index of the gate.
Input