Remote control commands
R&S
®
RTO6
1294User Manual 1801.6687.02 ─ 05
FORMat:BPATtern <BitPatternFormat>
Sets the number format for remote bit pattern queries on serial protocols.
Parameters:
<BitPatternFormat> DEC | HEX | OCT | BIN | ASCII | ASCii | STRG
ASCII = ASCii
*RST: HEX
Usage: Asynchronous command
SYSTem:DISPlay:UPDate <Enable>
Defines whether the display is updated while the instrument is in the remote state. If
the display is switched off, the normal GUI is replaced by a static image while the
instrument is in the remote state. Switching of
f the display can speed up the measure-
ment. This is the recommended state.
See also: Chapter 22.5.3.2, "Using the display during remote control", on page 1245
Parameters:
<Enable> ON | 1:
Display is shown and updated during remote control
OFF | 0: Display shows static image during remote control
Example:
SYSTem:
DISPlay:UPDate 1
Switch on the display update.
SYSTem:KLOCk <Enable>
Locks or unlocks the local controls of the instrument. This includes the front panel
keys, the keyboard, or other local interfaces. except for the "V
iew" button on the dis-
play.
Parameters:
<Enable> ON | 1: Locks the local keys
OFF | 0: Keys are unlocked
Usage: SCPI confirmed
GPIB:ADDRess <Address>
Sets the GPIB address of the instrument. Changing the address has major effects on
the communication to the remote computer
.
Parameters:
<Address> Range: 0 to 30
Increment: 1
*RST: 20
Usage: Asynchronous command