EasyManua.ls Logo

R&S RTO6 - Page 1295

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTO6
1295User Manual 1801.6687.02 ─ 05
GPIB:TERMinator <T
erminator>
Specifies the symbol that is used as a terminator in GPIB communication.
Parameters:
<T
erminator> LFEoi | EOI
*RST: EOI
Usage: Asynchronous command
SYSTem:DISPlay:MESSage:STATe <DispMessState>
Enables and disables the display of an additonal text in remote control.
To define the text, use
SYSTem:DISPlay:MESSage[:TEXT].
Parameters:
<DispMessState> ON | OFF
*RST: OFF
Usage: Asynchronous command
Firmware/software: Version 2.70
SYSTem:DISPlay:MESSage[:TEXT] <DisplayMessage>
Defines an additional text that is displayed during remote control operation.
To enable the text display
, use SYSTem:DISPlay:MESSage:STATe on page 1295.
Parameters:
<DisplayMessage> String that contains the text.
Usage: Asynchronous command
Firmware/software: Version 2.70
SYSTem:LANGuage <Language>
Defines the remote control behavior of the instrument and sets the remote control com-
mand set.
Parameters:
<Language> String value. Available values:
'SCPI': R&S R
T
O6 remote command set is used.
"DPO7000" or 'TDS540': Compatible remote command set of
Tektronix oscilloscopes DPO7000 or TDS540 is used.
If one of these emulation modes is used, you can define alterna-
tive responses to the IDN*? and OPT*? commands on the
"Menu" > "Settings" > "System" > "Remote" > "SCPI Emulation"
tab.
Usage: Asynchronous command
General remote settings

Table of Contents

Other manuals for R&S RTO6

Related product manuals