Remote control commands
R&S
®
RTO6
1317User Manual 1801.6687.02 ─ 05
MTESt:LABel <LabelState>
Switches the display of the mask test name on or of
f.
T
o change the name of the mask test, use MTESt:REName on page 1317.
Parameters:
<LabelState> ON | OFF
*RST: ON
Usage: Asynchronous command
MTESt:REName <MaskTestName>, <NewName>
MTESt:REName
? <MaskTestName>
Changes the name of the mask test.
Parameters:
<NewName> String with the new mask test name.
Parameters for setting and query:
<MaskTestName> String with the existing mask test name.
MTESt:HIGHlight:INFinite <State>
If ON, the mask hits are highlighted for an unlimited period of time.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
MTESt:HIGHlight:STATe <HighlightHits>
If ON, the mask hits are highlighted on the screen. You can define the color and the
time of the hit display
.
Parameters:
<HighlightHits> ON | OFF
*RST: ON
Usage: Asynchronous command
MTESt:HIGHlight:TIME <HighlightTime>
Sets the time how long the mask hits are highlighted.
Instrument setup