Remote control commands
R&S
®
RTO6
1355User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Parameters:
<MeasMode> DMODe | CMODe | PMODe | NMODe
DMODe
Differential mode input voltage (V
dm
), the voltage between the
positive and negative input terminal.
CMODe
Common mode input voltage (V
cm
) , the mean voltage between
the positive and negative input terminal vs. ground.
PMODe
Positive single-ended input voltage (V
p
). the voltage between the
positive input terminal and ground.
NMODe
Negative single-ended input voltage (V
N
). the voltage between
the negative input terminal and ground.
*RST
: DMODe
Usage: Asynchronous command
PROBe<m>:SETup:DMOFfset <DMOf
fset>
Sets the dif
ferential offset to compensate a DC voltage applied to the positive and the
negative input terminal.
Suffix:
<m>
.
1..4
Parameters:
<DMOffset> In "DM" probe mode (DMODe), this of
fset is used as channel of
f-
set and considered automatically for correction.
Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
PROBe<m>:SETup:CMOFfset <CMOffset>
Sets the common-mode of
fset. The setting is only available for differential probes.
Suffix:
<m>
.
1..4
Selects the input channel.
Acquisition and setup