Remote control commands
R&S
®
RTO6
1356User Manual 1801.6687.02 ─ 05
Parameters:
<CMOffset> Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
PROBe<m>:SETup:NOFFset <NOffset>
Sets the negative of
fset to compensate a DC voltage applied to the negative input ter-
minal (Vp) referenced to ground.
Suffix:
<m>
.
1..4
Parameters:
<NOffset> In "N" probe mode (NMODe), this of
fset is used as channel of
fset
and considered automatically for correction.
Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
PROBe<m>:SETup:POFFset <POffset>
Sets the positive of
fset to compensate a DC voltage applied to the positive input termi-
nal (Vp) referenced to ground.
Suffix:
<m>
.
1..4
Parameters:
<POffset> In "P" probe mode (PMODe), this of
fset is used as channel of
fset
and considered automatically for correction.
Range: -100E+24 to 100E+24
Increment: 1E-3
*RST: 0
Default unit: V
Usage: Asynchronous command
PROBe<m>:SETup:TERM:STATe <VoltageState>
Activates the instrument control of the termination voltage.
Suffix:
<m>
.
1..4
Acquisition and setup