TRIGger<m>:SPATtern:CSOurce[:VALue].......................................................................
141
1
TRIGger<m>:SPATtern:CSOurce:EDGE.........................................................................1411
TRIGger<m>:SPATtern:CSOurce:LEVel..........................................................................1411
TRIGger<m>:SPATtern:PATTern....................................................................................1412
TRIGger<m>:SPATtern:REFSource............................................................................... 1412
TRIGger<m>:SPATtern:CSOurce[:VALue] <ClockSource>
Defines the source of the clock signal.
Suffix:
<m>
.
1..3
Only 1 = A-trigger
, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockSource>
CHAN1 | CHANNEL1 | CHAN2 | CHANNEL2 | CHAN3 |
CHANNEL3 | CHAN4 | CHANNEL4
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
Input channel of the clock signal
*RST: CHAN1
Usage: Asynchronous command
TRIGger<m>:SPATtern:CSOurce:EDGE <ClockEdge>
Together with the clock level (see TRIGger<m>:
SPATtern:CSOurce:LEVel
on page 1411), the clock edge defines the point in time when the state of the data sig-
nal is checked.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockEdge>
POSitive | NEGative | EITHer
See Chapter 24.4.3, "Slope parameter", on page 1285.
*RST
: POSitive
Usage: Asynchronous command
TRIGger<m>:SPATtern:CSOurce:LEVel <ClockLevel>
Defines the voltage level for the clock signal.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.